For full functionality of this site it is necessary to enable JavaScript.
EMIN.MY
0
Product image

Trotec PC220 Particle Counter (0 ~ 2000 μg/m³)

ModelPC220
P/N3.510.006.015
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Dimensions (H x W x D): 240 mm x 75 mm x 57 mm

Weight: 570 g

Light source: Laser class 3R, wavelength 780 nm, 90 mW according to EN

PC interfaces: USB port

Tripod thread: 1/4 inch – 20 UNC

Storage conditions: -10 °C to +60 °C at 10 to90 % RH (non-condensing)

Operating conditions: 0 °C to +50 °C at 10 to 90 % RH (non-condensing)

Display: 2.8 inch colour LCD display with background illumination, 320 x 240 pixels

Functions: Maximum, minimum and average value display, measured value hold function, alarm function, language selection, °C/°F switch, photo or video recording

Image format, resolution: JPEG, 640 x 480 pixels

Video format, resolution: 3GP, 320 x 240 pixels

Data storage: 5000 data records on the internal flash memory (optional memory expansion with a microSD card: up to 16 GB)

Energy

Battery: Polymer LI-ION battery

Operating time: approx. 4 hours of continuous operation

Charging time: approx. 2 hours with an alternating current adapter

Automatic switch-off: 3 minutes, 15 minutes or 60 minutes

Automatic screen switch-off: 90 seconds, 2 minutes or 4 minutes

Temperature measurement

Temperature range: 0°C to 50°C (32°F to 122°F)

Temperature accuracy:

±0.5 °C (0.9 °F) at 10 °C to 40 °C (50 °F to 104 °F)

±1.0 °C (1.8 °F) in other temperature ranges

Dew point temperature range: 0°C to 50°C (32°F to 122°F)

Dew point temperature accuracy:

±0.5 °C (0.9 °F) at 10 °C to 40 °C (50 °F to 104 °F) ±1.0 °C (1.8 °F) at other temperature ranges

Wet-bulb temperature range: 0°C to 80°C (32°F to 176°F)

Wet-bulb temperature accuracy: ±1.0 °C (1.8 °F)

Humidity measurement

Humidity level measuring range: 0 % RH to 100 % RH

Humidity level accuracy:

±3 % RH at 40 % to 60 %

±3.5 % RH at 20 % to 40 % and 60 % to 80 %

±5 % RH at 0 % to 20 % and 80 % to 100 %

Particle counter

Channels (detectable particle sizes): 0.3 μm, 0.5 μm, 1.0 μm, 2.5 μm, 5.0 μm, 10.0 μm

Flow rate : 2.83 l/min. (0.1 ft³/min.) (=>0.99 l/21 s) controlled by internal pump

Counter mode: Cumulative, differential, concentration

Counting efficiency: 50 % at 0.3 μm; 100 % for particles > 0.45 μm (according to ISO 21501)

Coincidence loss: 5 %, 2 million particles per 28.3 litres

Zero count: 1 count/5 minutes (according to JIS B9921)

Start delay: 1 to 100 seconds

Sample inlet: Isokinetic probe

Calibration: Using monodisperse latex particles (PSL particles; compliant with NIST)

Light source of the measuring cell: Laser class 1 (class 3R laser enclosed in a way to prevent tampering, 780 nm, 1.5-3 mW, classified according to DIN EN 60285-1 and the Technical Regulations of the Occupational Health and Safety; Ordinance on Artificial Optical Radiation (TROS))

Mass concentration

Channels: PM2.5 / PM10

Measuring range: 0 to 2000 μg/m³

Resolution: 1 μg/m³

Formaldehyde (HCHO)

Measuring range: 0.01 to 5.00 ppm

Accuracy: ±5 % FS

Resolution: 0.01 ppm

Carbon monoxide (CO)

Measuring range: 10 to 1000 ppm

Accuracy: ±5 % FS

Resolution: 1 ppm

Scope of delivery

• 1 x Particle counter PC220

• 1 x Mini tripod

• 1 x USB connecting cable + software

• 1 x Quick guide

• 1 x Transport case

• 1 x Zero filter + connection hose

• 1 x Charger

Brochure

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support