For full functionality of this site it is necessary to enable JavaScript.
EMIN.MY
0
Product image

Sundy SDS-V Coulomb Sulfur Analyzer (0.01%–40%)

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Test Method: Coulomb Titration

Sulfur Determination Range: 0.01%–40%

Standards: ISO20336, GB/T 214-2007

Sulfur Resolution: 0.001%

Analysis Time/Sample: 3~6min

Preheating Time: ≤30min

Max. Quantity of Sample: 1

Sample Weight: 10–110mg (50±5mg Recommended)

Furnace Temperature: 1150℃ (coal), 900℃ (oil), 1190℃ (cement)

Heating Elements: Silicon Carbide

Temperature Control Precision: ±2.5℃

Temperature Sensor: Thermocouple

Power Requirement: AC 220V (-15%~10%) 50/60Hz

Max. Power: 3.5kW

Gas Requirement: Air

Gas Pressure: 14 to 45psig (1 to 3 bar)

Sample Introduction and Discarding: Manual

Gas Flow Checking: Rotor flow meter adopted, gas flow by visual check

Chemical Reagents: KI, KBr, Glacial acetic

Humidity: ≤85%

Ambient Temperature: ≤40℃

Operational Control: Windows®-based software on external PC

Operation Cost: Lower cost, Desiccant: allochroic silicagel

Dimensions: 680x520x395mm

Weight: 78kg

Advance Calculation: Yes

Analysis Report: Yes

Statistics Report: Yes

System Monitor: Yes

SDS-V Coulomb Sulfur Analyzer can be used to determine the total sulfur content in combustibles such as coal, coke, petroleum and other solid biomass fuels in the field of power plant, coal mining, cement & steel company, environmental industry.