Earth Continuity

Method: 2 wire technique, using 'zero' lead function.

Test Current: >+200mA, -200mA DC into 2 ohms.

Max Test Voltage: 4-24V RMS o/c.

Measuring Range (low range): 0.001 – 0.999 ohms @ 0.001 ohms resolution.

Measuring Range (mid range): 1.00 – 9.99 ohms @ 0.01 ohms resolution.

Measuring Range (high range): 10.0 – 19.9 ohms @ 0.1 ohms resolution.

Accuracy: ± 3% of reading + 10 m ohms.


Direct Leakage Measurement

Measuring Range: 4μA to 9999μA.

Accuracy: ± 5% of reading + 2 counts.

Measuring Device: As per IEC 60601-1 requirements.

Measurement Type: True RMS.


Alternative Leakage Measurements

Test Voltage: 250V at mains frequency.

Test Current: 3.5 mA current limited.

Measurement Range: 4μA to 9999μA.

Measurement Resolution: 1μA.

Measurement Accuracy: ±5% of reading + 2 counts.

Measurement Type: True RMS.

Measuring Device: As per IEC 60601-1.


Mains Outlet Test

Input Voltage Range: 0-300V AC.

Max Current: 16A.

Measures: L – E, N – E.

Accuracy: ± 5% of reading + 2 counts.


Insulation Resistance

Measurement: EUT to earth/ground, EUT to AP, AP to earth/ground.

Voltage: 50 / 100 / 250 & 500V DC @1mA.

Range (low range) @ 50V DC: 0.01Mohms - 10Mohms.

Range (low range) @ Above 50V DC: 0.01Mohms - 20Mohms.

Accuracy (low range): ± 5% of reading + 2 counts.

Range (high range) @ 250V DC: 20Mohms – 50Mohms.

Range (high range) @ 500V DC: 20Mohms – 100Mohms.

Accuracy (high range): ±10% + 2 counts.

Resolution: 0.01Mohms.


Differential Leakage Measurement

Measuring Range: 75μA to 9999μA.

Accuracy: ±5% of reading + 5 counts.

Measurement / Display Resolution: 1μA.

Measurement Type: True RMS.

Measuring Device: Frequency response characteristics to IEC 60601-1.


Power Measurement

Method: VA rating.

Range: 0.1kVA – 4kVA.

Accuracy: ±10% + 2 counts.


IEC Mains Lead Test

Test Duration: IEC Mains Lead Test.

Test: Continuity of all conductors, Insulation, and Polarity.


General Specification

Mains Power: 230V AC ±10%, 50-60Hz ± 1Hz

Battery: 6 x 1.5V AA (Batteries not included).

Weights: 1.6 kg / 3.5 lbs including batteries.

Size (L x W x D): 270 x 110 x 75 mm / 10.5 x 4 x 3”.