PMS SLS-1100 Syringe Sampling System with LiQuilaz® Particle Counter (UltraChem 100)

ModelPMS SLS-1100
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

+ Small volume chemical quality assurance
+ Parts cleanliness testing
+ Laboratory water sampling
+ Pharmaceutical testing for USP compliance
+ Industries: semiconductor, pharmaceutical, data storage, medical, aerospace, automotive, hydraulic oils

Syringe size: 25 ml
Sampling mode: Vacuum, Batch
Minimum sample size: 0.4 ml, 20 ml/min
Wetted surface materials: Glass, Teflon® (all Teflon optional)
Power: 100 – 240 V, 50 – 60Hz
Sample volume: Adjustable using software interface
Dimensions (l, w, h): 38.1 x 22.0 x 43.4 cm, including particle counter
Weight: 19.1 kg including particle counter
Environment Temperature: 10 – 30 ºC, humidity: non-condensing
Particle Sensor: UltraChem 100
Particle Size range: 100 – 500 nm
Channels: 100, 200, 300, 500 nm
Flow rate: 20 ml/min
Sampled %: 30% Monitor mode; 16% Spectrometer mode, 13% High-Scatter mode
Maximum concentration: 3000 particles/ml Monitor mode, 9000 particles/ml Spectrometer mode, 15000 particles /ml, High-Scatter mode
Zero count: < 50 counts/l
Sample temperature: 50 – 122 °F (10 – 50 ºC)
Calibration: Materials used are traceable to US National Institute of Standards and Technology (NIST) and Japanese
Institute of Standards (JIS).
Software: SamplerSight