Particles Plus 9300P-OEM Airborne Particle Counter Sensor (0.3 - 25 μm; 1.2 LPM / ±3%)

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Size Range: 0.3 to 25 μm

Size Channels: Factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm

Counting Efficiency: 50% @ 0.3μm; 100% for particles > 0.45 μm per JIS

Flow Rate / Accuracy: 1.2 LPM / ±3%

Concentration Limits: 27,000,000 particles/ft3 @ 10% coincidence (per ISO 21501-4), No appreciable drift. 50,000,000 particles/ft3 @ 10% coincidence (as tested and validated1)

Aerosol Concentration Range: 0.01 to 20,000 μg/m3

Light Source: Long life laser diode

Zero Count: <1 count / 60 minutes (<1 particles / 6 ft3). No fault count subtraction. No appreciable drift.

Resolution: 0.00003 μg/m3

Calibration: NIST Traceable

Internal Vacuum Pump: Integrated pump

Filtered Exhaust: HEPA filtered exhaust

Airflow: Internally monitored

Number of Channels: Up to 30 channels

Alarm: Adjustable built-in alarm

Communication Modes: Modbus (RTU/ASCII) via 3.3v UART

Standards: ISO 21501-4 and JIS B9921

Instrument Calibration: Recommended minimum once a year

Dimensions (L x W x H): 3.41 x 4.89 x 1.57 in (8.7 x 12.4 x 4.0 cm) includes inlet

Weight: 13 oz (368 grams)

Optional Accessories: 9300P-OEM Starter kit, 9000 Series Breakout Board, Square Base with flanges

Data Storage: 65,000 sample records (rotating buffer) with environmental data

Sample Time: 1 second to 99 hours

Power: 7 - 15 Vdc (< 2 watts max)

Operating Conditions: 41º to 104ºF (5º to 40ºC) / 20% to 95% non-condensing RH

Storage Conditions: 32° to 122°F (0° - 50°C) / < 98% non-condensing RH

Datasheet

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support