Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
pH Range: -2.00 to 16.00 pH / -2.000 to 16.000 pH
mV Range: ±699.9 mV / ±2000 mV
EC Range: 0.00 to 29.99 µS/cm / 30.0 to 299.9 µS/cm
EC Range: 300 to 2999 µS/cm
EC Range: 3.00 to 29.99 mS/cm / 30.0 to 200.0 mS/cm
EC Range: up to 500.0 mS/cm (uncompensated EC)*
TDS Range: 0.00 to 14.99 ppm (mg/L)
TDS Range: 15.0 to 149.9 ppm (mg/L) / 150 to 1499 ppm (mg/L)
TDS Range: 1.50 to 14.99 ppt (g/L) / 15.0 to 100.0 ppt (g/L)
TDS Range: up to 400 g/L (uncompensated TDS)* (with 0.80 factor)
NaCl Range: 0.0 to 400.0%
Temp. Range: -20.0 to 120.00°C (-4.0 to 248.0°F)
pH Resolution: 0.01 pH / 0.001 pH
mV Resolution: 0.1 mV / 1 mV
EC Resolution: 0.01 uS/cm / 0.1 uS/cm / 1 uS/cm
EC Resolution: 0.01 mS/cm / 0.1 mS/cm
TDS Resolution: 0.01 ppm (mg/L) / 0.1 ppm (mg/L) / 1 ppm (mg/L)
TDS Resolution: 0.01 ppt (g/L) / 0.1 ppt ( g/L)
NaCl Resolution: 0.1%
Temperature Resolution: 0.1°C / (0.1°F)
Accuracy @ 20°C / 68°F
pH: ±0.01 / ±0.002 pH
mV: ±0.2 mV / ±1 mV
EC: ±1% of reading / ±0.05 uS/cm or 1 digit, whichever is greater
TDS: ±1% of reading / ±0.03 ppm or 1 digit, whichever is greater
NaCl: ±1% of reading
Temperature: ±0.4°C (±0.8°F)
Rel mV offset: ±2000 mV
Calibration
pH: 1, 2 or 3 calibration points, with 7 memorized buffers: 1.68, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45
EC/TDS: 1 point slope calibration, 6 standards: 84 uS/cm, 1413 uS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm / 1 point offset calibration: 0.00 uS/cm
NaCl: 1 point, with MA9066 salinity calibration solution
Temperature: 2 points, 0 and 50°C (32 and 122°F)
Temperature Compensation: Automatic or manual from -20.0 to 120.0°C (-4.0 to 248.0°F)
Conductivity temperature coefficient: Selectable 0.00 to 6.00 % / °C (EC and TDS only)
TDS factor: 0.40 to 0.80 / Default value: 0.50
pH Probe: MA917B/1 (included)
Temperature Probe: MA831R (included)
EC Probe: MA814DB/1 (4 ring, included)
Logging-On-Demand: Maximum 50 samples for each range: (pH, mV /Rel mV, EC, TDS, NaCl)
PC Connectivity: RS232/USB opto isolated
Power Supply: 12 VDC adapter (included)
Dimensions: 230 x 160 x 95 mm (9.0 x 6.3 x 3.7 inches)
Environment: 0 to 50°C / 32 to 122°F; max RH 95%
Weight: 0.9 kg / 2.00 lbs.
*Uncompensated conductivity (TDS) is the conductivity (or TDS) value without temperature compensation.