Mettler Toledo 30034264 CO2 Sensor InPro5500i/D28/M42

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Accuracy: ± 1 % (± 5 °C of calibration temperature, ± 2 % over temperature range 0 … 50 °C (32 … 122 °F)

IP Rating: IP 67

Model Family: InPro 5500i

CIP resistant: Yes

Response Time: 21 s (measurement cycle time)

Cable Connector: M12 5 PIN

Design Pressure: 290 psi (20 bar)

O-Ring Material: EPDM

Connector Design: Straight

Flow Requirements: >0.5 m/s

Measurement Cycle: < 20 s

Process Connection: 29 mm with cap nut M42

Operating Range CO2: 0 psi – 145 psi (0 bar – 10 bar)

Measurement Parameter: Carbon Dioxide

Measuring Temp. Range: 32 °F – 122 °F (0 °C – 50 °C)

Measurement Technology: Thermal Conductivity

Mechanical Temp. Range: 23 °F – 250 °F (-5 °C – 121 °C)

ISM (Digital) or Analog: ISM (Digital)

Certificates & Approvals: Quality certificate, Material certificate 3.1, Surface finish certificate

Operating Pressure (Bar): 0-20 bar

Operating Pressure (PSI): 362.6 psi

Temperature compensation: Automatic with built -in RTD

Wetted Metallic Material: 316L stainless steel

Wetted membrane material: Silicone

Autoclavable/Sterilizable: Sterilizable Only

Operating Range CO2 (g/L): 0 – 15

Primary Media/Application: Food and Beverage

Operational Range CO2 (V/V): 0 – 7

Measurement Range: Concentration: 0…10 bar pCO2, 0…15 g/L CO2, 0...7 V/V CO2

Measurement Range (Concentration): 0…145 psig pCO2

Hygienic/Biocompatibility Approvals: FDA

Surface Roughnes of Wetted Material: N5/Ra16 (Ra=0.4μm/16μin)