JEIOTECH HL-45H Recirculating Coolers (-20~40℃, 2hp)

ModelHL-45H
P/NAAH65168K
USD 11,939.40
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Temperature Range(℃ / ℉) -20 to 40 / -4 to 104
Temperature Stabilityat 15℃, (±℃ / ℉) 1 / 1.8
Cooling Capacity: at 20℃ (kw)-0.6/ at 10℃ (kw) - 6.5/ at 0℃ (kw) - 4.5
at -10℃ (kw) - 2.1/ at -20℃ (kw) - 1.2
Pump: Max. Flow Rate (L / min, gal / min) 70 lit
: Max. Pressure (bar / psi) 6 bar
Dimension: Overall (WxLxH, mm / inch) 605×1045×1300 mm
Net Weight (kg / Ibs) 171 kg
Electric Requirement: 380V, 50Hz

Operating features


• Temperature range is from -20℃~40℃.
• Bright LED water level indicator can be easily identified.
• Low fluid level and dry-running protection when triggered shows an error symbol on the display and an 
   audible sound.
• If the equipment operates at 40℃ or above due to the external environment, user can simply identify the 
   error with a visual or an audible sound.
• Designed to simplify operation with a touch screen type display.
  - Easy-to-read VFD display with interactive touch keys.
  - Setting and resolution indication 0.1℃/℉.
  - Quick keypad lock prevents accidental parameter changes.
  - Signal indicator for operation status.


Constructional features


• The refrigerating system is designed to reduce energy consumption and to maximize in refrigerating 
   efficiency.
• Eco-friendly R-404A / R507 refrigerant is used to prevent environmental pollution. Double service valve is 
   used to prevent refrigerant leakage.
• Removable gridded vent for reducing condenser contamination and to facilitate cleanliness.
• A wide (Ø120mm / 4.72″) solution inlet is for easy and safe pouring of solution into the recirculating cooler.
• Pressure gauge comes standard for observing the pump pressure
• The refrigeration system can be observed easily to diagnostic at a glance by high / low pressure gauges. 
   (HL-25H, 35H, 45H, 55H models)
• Wheels allow for easy mobility.
• The beautifully designed and compact.
• Different types of pumps come standard with the recirculating cooler to provide versatility.
• Warning alarm/monitoring system.
  - Over temperature.
  - Lack of bath solutions.
• Variety of accessories is available to meet user’s different lab environment needs. (Refer to accessories)

Model HL
-05
HL
-10
HL
-15
HL
-20
HL
-15H
HL
-20H
HL
-25H
HL
-35H
HL
-45H2)
HL
-55H2)
Temperature1)
(bath fluid :
Ethtylene glycol mix
1:1 Water)
Working Temperature
range 
(℃ / ℉)
-20 to 40 / -4 to 104
Temperature stability
at 15℃, (±℃ / ℉)
1 / 1.8
Cooling capacity
(bath fluid : Ethtylene
glycol mix 1:1 Water)
at 20℃ (kw)
0.60
0.07
1.45
1.80
1.45
1.80
2.50
3.30
6.50
7.10
at 10℃ (kw)
0.50
0.60
1.15
1.50
1.15
1.50
1.80
2.30
4.50
6.00
at 0℃ (kw)
0.35
0.42
0.86
1.15
0.86
1.15
1.10
1.70
3.00
4.10
at -10℃ (kw)
0.18
0.32
0.62
0.85
0.62
0.85
0.65
1.00
2.10
2.50
at -20℃ (kw)
0.04
0.14
0.30
0.40
0.30
0.40
0.40
0.55
1.20
1.50
Pump Max. flow rate
(L / min, gal / min)
50 / 13.2
40 / 10.57
70 / 18.49
Max. pressure
(bar / psi)
1.0 / 14.5
2.8 / 40.6
6 / 87.0
Dimension Max,. filling capacity
(L, cu ft)
10 / 0.35
15 / 0.53
15 / 0.53
25 / 0.88
40 /1.41
For tubing dia
(mm / inch)
12.7 / 0.5
20 / 0.79
Overall
(WxLxH, mm / inch)
405×620×710 /
15.9×24.4×28.0
515×715× 835 /
20.3×28.1×32.9
515×715× 835 /
20.3×28.1×32.9
550×900×1140 /
21.7×35.4×44.9
605×1045×1300 /
23.8×41.4×51.2
Net weight (kg / Ibs)
62.85 /
138.6
64.85/
143.0
86.5 /
190.7
87.35 /
192.9
91.8 /
202.4
92.65 /
204.3
141.3 /
310.9
146.3 /
321.9
171 ±10 /
377 ±22
176 ±10 /
388 ±10
Electric requirement 
(230V, 60Hz)
5.0A
5.5A
6.5A
7.5A
6.5A
7.5A
9.5A
13.5A
   
Cat. No.
AAH
65001
AAH
65011
AAH
65021
AAH
65031
AAH
65121
AAH
65131
AAH
65141
AAH
65151
-
Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts