HANNA HI98326 Dist9 Salt/Temperature Pen (70.0 ppt / 0.1 ppt; 50.0°C / 0.1°C)

ModelHI98326
OriginRumani
USD 113.44
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

SALINITY ACCORDING TO PPT (G/L)

LOW LADDER

Scale: 0.00 to 10.00 ppt (g/L)

Resolution: 0.01 ppt (g/L)

Accuracy: ±0.20 ppt (g/L)

HIGH LADDER

Scale: 0.0 to 70.0 ppt (g/L)

Resolution: 0.1 ppt (g/L)

Accuracy:

±1.0 ppt (g/L) (scale from 0.0 to 40.0)

±2.0 ppt (g/L) (range 40.0 to 70.0)

AUTO (DEFAULT)

The scale:

0.00 to 9.99 ppt (g/L)

10.0 to 70.0 ppt (g/L)

Resolution: 0.01 / 0.1 ppt (g/L)

Accuracy:

±0.20 ppt (g/L) (scale from 0.00 to 9.99)

±1.0 ppt (g/L) (scale from 0.0 to 40.0)

±2.0 ppt (g/L) (range 40.0 to 70.0)

SALINITY BY PSU

LOW LADDER

Scale: 0.00 to 10.00 PSU

Resolution: 0.01 PSU

Accuracy: ±0.20 PSU

HIGH LADDER

Scale: 0.0 to 70.0 PSU

Resolution: 0.1 PSU

Accuracy

±1.0 PSU (scale from 0.0 to 40.0)

±2.0 PSU (range from 40.0 to 70.0)

AUTO (DEFAULT)

The scale:

0.00 to 9.99 PSU

10.0 to 70.0 PSUs

Resolution: 0.01 / 0.1 PSU

Accuracy

±0.20 PSU (scale from 0.00 to 9.99)

±1.0 PSU (scale from 0.0 to 40.0)

±2.0 PSU (range from 40.0 to 70.0)

SALINITY ACCORDING TO S.G.

LOW LADDER

Scale: 1.000 to 1.007 S.G.

Resolution: 0.001 S.G.

Accuracy: ±0.001 S.G.


HIGH LADDER

Scale: 1,000 to 1,041 S.G.

Resolution: 0.001 S.G.


Accuracy: ±0.001 S.G.

AUTO (DEFAULT)

Scale: 1,000 to 1,041 S.G.

Resolution: 0.001 S.G.

Accuracy: ±0.001 S.G.


TEMPERATURE

Range: 0.0 to 50.0 oC (32.0 to 122.0 oF)

Resolution: 0.1 oC / 0.1 oF

Accuracy: ±0.5 oC / ±0.1 oF

Method

ppt: International Oceanographic Tables, 1966

PSU: Standard Methods for the Examination of Water and

Wastewater, 2520 B, Electrical Conductivity Method

S.G.: Standard Methods for the Examination of Water and

Wastewater, 2520 C, Density Method


Calibration: Automatic, 1 or 2 points at 5.00 ppt or 35.00 ppt

Temperature compensation: Automatic from 5.0 to 50.0 oC (41.0 to 122.0 oF)

Battery: 1 CR2032 3V Lithium-ion battery

Can be used for about 100 hours continuously

Automatically turns off

Customization: After 8 minutes, 60 minutes

Turn off the auto-off feature

Environment: 0 to 50 °C (32 °C to 122 °F); RH max 100%


Supply includes

- HI98326 meter

- Battery (in the device)

- Calibration solution 5.00 ppt (2 packs) and 35.00 ppt (2 packs)

- Protective cap

- User manual

- Plastic container

 

Stay Updated with Offers