For full functionality of this site it is necessary to enable JavaScript.
EMIN.MY
0
Product image

Gastec 76H Quick-measuring Detector tubes Tetrahydrothiophene C4H8S

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Indicator: 10 to 200 ppm
Measuring Range: 10 to 200 ppm, Number of Pump Strokes 1
Sampling Time: 1 minute per pump stroke
Colour Change: Pink → Pale Yellow
Reaction Principle: C4H8S + MnO4 (Pink) → Pale Yellow Product
Coefficient of Variation: 10% (for 10 to 60 ppm), 5% (for 60 to 200 ppm)
Nr. of tests: 5
Pump: GV-100S/110S

Catalog

Manual

Performance
Measuring Range 10 to 200 ppm
Number of Pump Strokes 1
Correction Factor 1
Sampling Time 1 minute per pump stroke
Detecting Limit 2 ppm (n=1)
Colour Change Pink → Pale Yellow
Reaction Principle C4H8S + MnO4 (Pink) → Pale Yellow Product
Coefficient of Variation 10% (for 10 to 60 ppm), 5% (for 60 to 200 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Ethyl mercaptan 3 ppm No Pale pink (whole layer)
Ethylene 5000 ppm No Pale pink (whole layer)
Propylene 2000 ppm No Pale pink (whole layer)
Methanol 2000 ppm No Pale pink (whole layer)
Hydrogen sulphide 300 ppm No Pale pink (whole layer)
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
- - -

 