Gastec 6 Quick-measuring Detector tubes Water vapour H2O (0.5~32mg/l)
ManufacturerGastec(View more products from this manufacturer)
Model6
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available
Indicator: 1 to 18 mg/L
Measuring Range:
+ 0.5 to 1 mg/L, Number of Pump Strokes 2
+ 1 to 18 mg/L, Number of Pump Strokes 1
+ 18 to 32 mg/L, Number of Pump Strokes 1/2
Sampling Time: 45 seconds per pump stroke
Colour Change: Yellowish Green → Purple
Reaction Principle: H2O + Mg(ClO4)2 → Mg(ClO4)2 H2O
Coefficient of Variation: 10% (for 1 to 6 mg/L), 5% (for 6 to 18 mg/L)
Nr. of tests: 10
Pump: GV-100S/110S
Performance
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 45 seconds per pump stroke | |||
| Detecting Limit | 0.1 mg/L (n=2) | |||
| Colour Change | Yellowish Green → Purple | |||
| Reaction Principle | H2O + Mg(ClO4)2 → Mg(ClO4)2 |
|||
| Coefficient of Variation | 10% (for 1 to 6 mg/L), 5% (for 6 to 18 mg/L) | |||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity | Temperature correction is necessary | |||
| Store the tubes at cool and dark place. | ||||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Alcohols, Ketones, Aldehydes, sulphur dioxide, Hydrogen chloride | ≧ 2000 ppm | + | No |
| Esters, Acrylonitrile | ≧ 5000 ppm | + | No |
| Ammonia | ≧ 1000 ppm | + | No |
| Nitrogen dioxide, Hydrogen cyanide |
≧ 500 ppm | + | No |
| Hydrogen sulphide | No | No |
| Calibration gas generation | Vaper pressure method |
| TLV-TWA | TLV-STEL | Explosive range |
| - | - | - |
Stay Updated with Offers
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Quick Support