Gastec 4LL Quick-measuring Detector tubes Hydrogen sulphide H2S (0.25-120 ppm)
ManufacturerGastec(View more products from this manufacturer)
Model4LL
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available
Indicator: 5 to 60 ppm
Measuring Range:
+ 0.25 to 2.5 ppm, Number of Pump Strokes 2 to 10
+ 2.5 to 60 ppm, Number of Pump Strokes 1
+ 60 to 120 ppm, Number of Pump Strokes 1/2
Sampling Time: per pump stroke
Colour Change: White → Brown
Reaction Principle: H2S + Pb(CH3COO)2 → PbS + 2CH3COOH
Coefficient of Variation: 5% (for 2.5 to 60 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 1 minute per pump stroke | |||
| Detecting Limit | 0.1 ppm (n=10) | |||
| Colour Change | White → Brown | |||
| Reaction Principle | H2S + Pb(CH3COO)2 → PbS + 2CH3COOH | |||
| Coefficient of Variation | 5% (for 2.5 to 60 ppm) | |||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity | Unnecessary | |||
| Store the tubes at cool and dark place. | ||||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Nitrogen dioxide | 1/4 time | No | |
| Sulphur dioxide | 1/1 | + | No |
| Mercaptans, Hydroren cyanide |
No | No | |
| Ozone | Bleaching | No |
| Calibration gas generation | Permeation tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 1 ppm (2014) | 5 ppm (2014) | 4.0 to 44% |