Gastec 21LA Quick-measuring Detector tubes Carbonyl sulphide COS (2-125ppm)
ManufacturerGastec(View more products from this manufacturer)
Model21LA
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available
Indicator: 5 to 50 ppm
Measuring Range:
+ 2 to 5 ppm, Number of Pump Strokes 2
+ 5 to 50 ppm, Number of Pump Strokes 1
+ 50 to 125 ppm, Number of Pump Strokes 1/2
Sampling Time: 3 minutes per pump stroke
Colour Change: Bluish Purple → White
Reaction Principle: COS + I2O5 + H2SO4 → SO2 + CO2, SO2 + I2 + H2O → 2HI + H2SO4
Coefficient of Variation: 10% (for 5 to 10 ppm), 5% (for 10 to 50 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
| +(Black) | Black at 500ppm | ||
| sulphur dioxide, Carbon disulphide | + | White | |
| Carbon dioxide | No | No | |
| Nitrogen dioxide | + | Brown at 2ppm | |
| Hydrogen sulphide | No effect at 1 pump stroke | *1 | |
| Propane, Butane | No | Black at the top of the discoloured area. | |
*1: If the primary tube is totally discoloured, the analyzer tube will be discoloured white due to the effect of hydrogen sulphide. |
|||
| Calibration gas generation | High pressure gas cylinder method |
| TLV-TWA | TLV-STEL | Explosive range |
| 5ppm(2014) | - | 12 to 19% |