Gastec 165L Quick-measuring Detector tubes Ethylene glycol HOCH2CH2OH (10 ~ 100 mg/m3)
ManufacturerGastec(View more products from this manufacturer)
Model165L
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available
Indicator: 10 ~ 100 mg/m3
Measuring Range: 10 ~ 100 mg/m3, Number of Pump Strokes 2
Sampling Time: 3 minute per pump stroke
Colour Change: Yellow → Reddish brown
Reaction Principle:
+ HOCH2CH2OH + Oxidiser → 2HCHO
+ 3HCHO + (NH2OH)3H3PO4 → H3PO4
+ H3PO4 + Base → Phosphate
Coefficient of Variation: 15% (for 10 to 20 mg/m3),10% (for 20 to 100 mg/m3)
Nr. of tests: 5
Pump: GV-100S/110S
Performance
| Measuring Range | 10 to 100 mg/m3 |
| Number of Pump Strokes | 2 |
| Correction Factor | 1 |
| Sampling Time | 3 minutes per pump stroke |
| Detecting Limit | 2 mg/m3 (n=2) |
| Colour Change | Yellow → Reddish brown |
| Reaction Principle | HOCH2CH2OH + Oxidiser → 2HCHO 3HCHO + (NH2OH)3H3PO4 → H3PO4 H3PO4 + Base → Phosphate |
| Coefficient of Variation | 15% (for 10 to 20 mg/m3),10% (for 20 to 100 mg/m3) |
| Shelf Life | 3 Years |
| Corrections for temperature & humidity | Temperature correction is necessary |
| Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. | |
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Aldehydes, Ketones, Acid gases | + | Reddish brown | |
| Tetrahydrothiophene, tert-Butyl mercaptan | + | Reddish brown( |
| Calibration gas generation | Vaper pressure method |
| TLV-TWA | TLV-STEL | Explosive range |
| - | C 100 mg/m3(H)(2014) | 3.2 to 33% |
Stay Updated with Offers
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Quick Support
Direct access to our certified experts