Gastec 16 Quick-measuring Detector tubes Phosgene COCl2 (0.05~20ppm)
ManufacturerGastec(View more products from this manufacturer)
Model16
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available
Indicator: 0.1 to 5 ppm
Measuring Range:
+ 0.05 to 0.1 ppm, Number of Pump Strokes 10
+ 0.1 to 5 ppm, Number of Pump Strokes 5
+ 5 to 20 ppm, Number of Pump Strokes 1
Sampling Time: per pump stroke
Colour Change: White → Yellow
Reaction Principle: COCl2+ (CH3)2NC6H4CHO → (CH3)2NC6H4CHCl2 + CO2, (CH3)2NC6H4CHCl2 + (C6H5)2NH → Yellow Product
Coefficient of Variation: 10% (for 0.1 to 1 ppm), 5% (for 1 to 5 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Performance
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 1 minute per pump stroke | |||
| Detecting Limit | 0.01 ppm (n=10) | |||
| Colour Change | White → Yellow | |||
| Reaction Principle | COCl2+ (CH3)2NC6H4CHO → (CH3)2NC6H4CHCl2 + CO2 (CH3)2NC6H4CHCl2 + (C6H5)2NH → Yellow Product |
|||
| Coefficient of Variation | 10% (for 0.1 to 1 ppm), 5% (for 1 to 5 ppm) | |||
| Shelf Life | 1.5 Years | |||
| Corrections for temperature & humidity | Temperature correction is necessary | |||
| Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. | ||||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Hydrogen chloride | 1/10 time | + | Yellow |
| Chlorine | 1/2 time | + | Yellow |
| Nitrogen dioxide | 1/5 time | + | Yellow |
| Calibration gas generation | Permeation tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 0.1 ppm (2014) | - |