Gastec 146 Quick-measuring Detector tubes Isopropyl acetate CH3CO2CH(CH3)2 (10 ~ 500 ppm)
ManufacturerGastec(View more products from this manufacturer)
Model146
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available
Indicator: 10 ~ 500 ppm
Measuring Range: 10 ~ 500 ppm, Number of Pump Strokes 2
Sampling Time: 2 minute per pump stroke
Colour Change: Yellow → Blackish brown → Pale blue after few minutes
Reaction Principle: CH3CO2CH(CH3)2+ Cr6+ + H2SO4 → Cr3+
Coefficient of Variation: 15% (for 20 to 100 ppm), 10% (for 100 to 500 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Performance
| Measuring Range | 10 to 500 ppm |
| Number of Pump Strokes | 2 |
| Correction Factor | 1 |
| Sampling Time | 2 minutes per pump stroke |
| Detecting Limit | 5 ppm (n=2) |
| Colour Change | Yellow → Blackish brown → Pale blue after few minutes |
| Reaction Principle | CH3CO2CH(CH3)2+ Cr6+ + H2SO4 → Cr3+ |
| Coefficient of Variation | 15% (for 20 to 100 ppm), 10% (for 100 to 500 ppm) |
| Shelf Life | 2 Years |
| Corrections for temperature & humidity | Temperature correction is necessary |
| Store the tubes at cool and dark place. | |
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Alcohols | + | Blackish brown then turns to pale blue | |
| Ketones | + | Blackish brown then turns to pale blue | |
| Esters | + | Blackish brown then turns to pale blue |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 100 ppm | 200 ppm | 1.8 to 8% |
Stay Updated with Offers
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Quick Support
Direct access to our certified experts

