Gastec 143 Quick-measuring Detector tubes Vinyl acetate CH3CO2CH:CH2 (5~250ppm)
ManufacturerGastec(View more products from this manufacturer)
Model143
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available
Indicator: 10 to 100 ppm
Measuring Range:
+ 5 to 10 ppm, Number of Pump Strokes 4
+ 10 to 100 ppm, Number of Pump Strokes 2
+ 100 to 250 ppm, Number of Pump Strokes 1
Sampling Time: 30 seconds per pump stroke
Colour Change: White → Brown
Reaction Principle: CH3CO2CH:CH2+C6H3(CH3)3+ H2S2O7 → Condensation Polymer (Brown)
Coefficient of Variation: 15% (for 10 to 20 ppm), 10% (for 20 to 100 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
Performance
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 30 seconds per pump stroke | |||
| Detecting Limit | 1 ppm (n=4) | |||
| Colour Change | White → Brown | |||
| Reaction Principle | CH3CO2CH:CH2+C6H3(CH3)3+ H2S2O7 → Condensation Polymer (Brown) | |||
| Coefficient of Variation | 15% (for 10 to 20 ppm), 10% (for 20 to 100 ppm) | |||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity | Unnecessary | |||
| Store the tubes at cool and dark place. | ||||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Aldehydes | + | Brown | |
| Ketones | + | Brown | |
| Ethylene | ≧ 1000 ppm | + | Yellow |
| Vinyl chloride | ≧ 500 ppm | + | Yellow |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 10 ppm | 15 ppm | 2.6 to 13.4% |