Gastec 135 Quick-measuring Detector tubes 1,1,1-Trichloroethane CH3CCl3 (100 ~ 2000 ppm)
ManufacturerGastec(View more products from this manufacturer)
Model135
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available
Indicator: 100 to 500 ppm
Measuring Range:
+ 100 to 500 ppm, Number of Pump Strokes 1
+ 500 to 2000 ppm, Number of Pump Strokes 1/2
Sampling Time: 3 minute per pump stroke
Colour Change: White → Reddish orange
Coefficient of Variation: 10% (for 100 to 200 ppm), 5% (for 200 to 500 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
Performance
| Measuring Range |
|
||
| Number of Pump Strokes |
|
||
| Correction Factor |
|
||
| Sampling Time | 3 minutes per pump stroke | ||
| Detecting Limit | 50 ppm (n=1) | ||
| Colour Change | White → Reddish orange | ||
| Reaction Principle | 1,1,1-Trichloroethane reacts with oxidizing agent to produce intermediate products then it react with detecting agent to produce reddish orange stain. | ||
| Coefficient of Variation | 10% (for 100 to 200 ppm), 5% (for 200 to 500 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity | Temperature correction is necessary | ||
| Store the tubes at cool and dark place. | |||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Chlorine, Bromine, Iodine | + | Reddish orange | |
| Chloroform, Dichloromethane | + | Reddish orange | |
| Carbon tetrachroride | No | No | |
| Methyl bromide | + | Reddish orange | |
| Trichloroethylene, Tetrachloroethylene | + | Reddish orange |
Other substance measurable with this detector tube
| Substance | Correction Factor | No. of pump strokes | Measuring Range |
| Chlorobromomethane | Factor: 0.22 | 1 | 22 to 110 ppm |
| 1,1-Dichloroethane | Factor: 0.9 | 1 | 90 to 4500 ppm |
| 1,2-Dichloroethane | Factor: 4.0 | 1 | 400 to 2000 ppm |
| 1,1,2-Trichloroethane | Tube 135 Reading (n=2) | 100 | 200 | 300 | 400 | 500 |
| 1,1,2-Trichloroethane (ppm) | 220 | 350 | 480 | 610 | 750 |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 350 ppm | 450 ppm | - |