For full functionality of this site it is necessary to enable JavaScript.
EMIN.MY
0
Product image

Gastec 123 Quick-measuring Detector tubes Xylene C6H4(CH3)2 (5~625ppm)

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Indicator: 20 to 250 ppm
Measuring Range:
+ 5 to10 ppm, Number of Pump Strokes 2
+ 10 to 250 ppm, Number of Pump Strokes 1
+ 250 to 625 ppm, Number of Pump Strokes 1/2
Sampling Time: 1.5 minute per pump stroke
Colour Change: White → Brown
Reaction Principle: C6H4(CH3)2 + I2O5 + H2SO4 → I2
Coefficient of Variation: 10% (for 10 to 50 ppm), 5% (for 50 to 250 ppm)
Nr. of tests: 10
Pump: GV-100S/110S

Catalog

Manual

Manual tube

Performance
Measuring Range
5 to10 ppm 10 to 250 ppm 250 to 625 ppm
Number of Pump Strokes
2 1 1/2
Correction Factor
1/2 1 2.5
Sampling Time
1.5 minutes per pump stroke 1 minute
Detecting Limit 1 ppm (n=2)
Colour Change White → Brown
Reaction Principle C6H4(CH3)2 + I2O5 + H2SO4 → I2
Coefficient of Variation 10% (for 10 to 50 ppm), 5% (for 50 to 250 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Carbon Monoxide ≦ 1000 ppm No(Two layers) Pale brown
(Whole layer)
Acetylene, Hexane 2000 ppm No(Two layers) Pale brown
(Whole layer)
Toluene ≧ 1/5 + Brown
Benzene ≧ 1/5 + Pale yellow
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Trimethylbenzene by scale 2 10 to 300ppm
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
100 ppm(2014) 150 ppm(2014) 1.0 to 7.0%

 

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts