Gastec 121SL Quick-measuring Detector tubes Benzene C6H6 (1~100ppm)
ManufacturerGastec(View more products from this manufacturer)
Model121SL
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available
Indicator: 1 to 20 ppm
Measuring Range:
+ 1 to 20 ppm, Number of Pump Strokes 5
+ 20 to 100 ppm, Number of Pump Strokes 1
Sampling Time: 2 minutes per pump stroke
Colour Change: White → Dark Green
Reaction Principle: C6H6 + I2O5 + H2S2O7 → I2
Coefficient of Variation: 20% (for 1 to 5 ppm), 15% (for 5 to 20 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
Performance
| Measuring Range |
|
||
| Number of Pump Strokes |
|
||
| Correction Factor |
|
||
| Sampling Time | 2 minutes per pump stroke | ||
| Detecting Limit | 0.25 ppm (n=5) | ||
| Colour Change | White → Dark Green | ||
| Reaction Principle | C6H6 + I2O5 + H2S2O7 → I2 | ||
| Coefficient of Variation | 20% (for 1 to 5 ppm), 15% (for 5 to 20 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity | Unnecessary | ||
| Store the tubes at cool and dark place. | |||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Hexane | 50 ppm | No | No |
| Toluene | 100 ppm | No | No |
| Xylene | 150 ppm | No | No |
Aromatic hydrocarbons other than benzene are trapped in the brown layer in the pretreatment tube. If the pretreatment reagent is entirely consumed (whole brown layer turns to dark brown), a higher reading will be given. |
|||
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 0.5 ppm(2014) | 2.5 ppm(2014) | 1.3 to 7.1% |