For full functionality of this site it is necessary to enable JavaScript.
EMIN.MY
0
Product image

Gastec 105 Quick-Measuring Detector Tubes Hydrocarbons (C5~C10 (100-3000Ppm))

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Indicator: 200 to 3000 ppm
Measuring Range:
+ 100 to 200 ppm, Number of Pump Strokes 2
+ 200 to 3000 ppm, Number of Pump Strokes 1
Sampling Time: 1.5 minute per pump stroke
Colour Change: White → Blackish brown
Reaction Principle: Hydorocarbon (C6 + C10) + I2O5 + H2S2O7 → I2
Coefficient of Variation: 10% (for 200 to 1000 ppm), 5% (for 1000 to 3000 ppm)
Nr. of tests: 10
Pump: GV-100S/110S

Catalog

Manual

Manual tube

Performance
Measuring Range
100 to 200 ppm200 to 3000 ppm
Number of Pump Strokes
21
Correction Factor
1/21
Sampling Time1.5 minutes per pump stroke
Detecting Limit20 ppm (n=2)
Colour ChangeWhite → Blackish brown
Reaction Principle

Hydorocarbon (C6 + C10) + I2O5 + H2S2O7 → I2

Coefficient of Variation10% (for 200 to 1000 ppm), 5% (for 1000 to 3000 ppm)
Shelf Life3 Years
Corrections for temperature & humidityUnnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
SubstanceConcentrationInterferenceChange colour by itself to
Acetylene, Ethylene0.1%+Blackish brown
Carbon monoxide0.1%+Blackish brown
Other organic solvents+Blackish brown
Other substance measurable with this detector tube
SubstanceCorrection FactorNo. of pump strokesMeasuring Range
Decane2.01 or 2200 to 6000 ppm
Heptane0.91 or 290 to 2700 ppm
Hexane0.81 or 280 to 2400 ppm
Nonane1.31 or 2130 to 3900 ppm
Octane1.01 or 21