Cosmopia ES-76EX Air Flow Type Thermal Shock Testing Apparatus (150℃/15 min; -65℃/15 min; 70L)

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Type: Stationary sample type, cool/hot air switchable, two/three zone switchable

Testing temperature range: Low-temperature testing: -70 to 0℃, High-temperature testing: 80 to 200℃

Temperature fluctuation margin: ±0.5℃

Temperature rise time (high-temperature chamber): Room temperature to 200℃: Within 25 min.

Temperature drop time (low-temperature chamber): Room temperature to -75℃: Within 70 min.

Temperature cycle performance:

- Cycles: Continuous operation cycles possible: 1,000

- Testing chamber temperature / Testing time (High): +150℃/15 min.

- Testing chamber temperature / Testing time (Low): -65℃/15 min.

- Temperature measurement position: Downwind side

- Sample Type: Plastic molded IC

- Sample Weight: 5 kg (IC 3.5 kg + single-stage jig 1.5 kg)

- Temperature return time: Within 5 min.

Permissible sample weight inside testing chamber: Max. 30 kg (floor surface 20 kg or less, sample shelf 5 kg or less [per shelf])

Testing chamber volume: 70 L

Testing chamber dimensions (W × D × H): 315 × 495 × 460 mm

Product dimensions (W × D × H): 1,565 × 1,640 × 1,980 mm

Exterior: Dark gray finish

Cooling method: Water-cooling, dual refrigeration, refrigerant: R404A (high-temperature side), R23 (low-temperature side)

Safeguards: Earth leakage circuit breaker, electric motor overload safeguard, electrical fuse, excess temperature increase prevention system, operation with door open prevention switch, temperature fuse, and high-voltage shutdown system for refrigerator

Accessories: Cable hole rubber plug × 1, sample shelf × 2, shelf support × 4, operation circuit protection fuse × 2, instruction manual × 1, and Y strainer × 2

Operational noise: 65 dB(A)

Dry air (Pressure): 0.4 to 0.7 MPa

Dry air (Consumption): Approx. 7.6 L(ANR) per cycle

Coolant: Volume: 4,200 L/h (coolant inlet temperature 32℃), Pressure: 0.1 to 0.5 MPa

Power supply: Three-phase 200V 50/60Hz, Three-phase 220V 60Hz, Three-phase 380V 50Hz

Maximum load power supply / [Power supply capacity]: 120 A / 42 kVA

Product weight: 1,600 kg