Type: Stationary sample type, cool/hot air switchable, two/three zone switchable

Testing temperature range: Low-temperature testing: -70 to 0℃, High-temperature testing: 60 to 200℃

Precooling temperature: -80℃ to testing chamber temperature low-temperature setting

Preheating temperature: Testing chamber temperature high-temperature setting to 210℃

Temperature fluctuation margin: ±0.5℃

Temperature rise time (high-temp chamber): Room temperature to 200℃: Within 20 min.

Temperature drop time (low-temp chamber): Room temperature to -75℃: Within 45 min.

Temperature return time: Within 10 min.

Return time guaranteed cycles: 5 cycles (low-temperature start)

Sample Type: Plastic molded IC

Sample Weight: 21 kg (IC 12 kg + dual-stage jig 9.0 kg)

Testing chamber temperature/time (Low): -65℃/30 min.

Testing chamber temperature/time (Room): Atmosphere temperature/10 min.

Testing chamber temperature/time (High): 150℃/30 min.

Temperature measurement position: Upwind side

Continuous operation cycles: 100 cycles (low-temperature start)

Testing chamber volume: 305L

Permissible sample weight inside testing chamber: Max. 50 kg (Bottom 40 kg or less, sample shelf 12 kg or less [per shelf])

Testing chamber dimensions (W×D×H): 720 × 920 × 460 mm

Product dimensions (W×D×H): 1,625 × 1,670 × 1,980 mm

System component materials: Exterior: Stainless steel (SUS430), Interior: Stainless steel (SUS304)

Cooling method: Air-Cooling Remote Condenser Type

Cooling system refrigerant/output (High side): R404A/7.4 kW (inverter control)

Cooling system refrigerant/output (Low side): R23/4.5 kW (inverter control)

Operational noise: 65 dB(A)

Accessories: Cable hole plug x1, shelf x2, support x4, fuse x2, manual x1, remote condenser x1

Power supply: Three-phase 200V 50/60Hz, 220V 60Hz, 380V 50Hz

Maximum load current / Power supply capacity: 125 A / 43 kVA

Dry air: Pressure: 0.4 to 0.7 MPa, Consumption: Approx. 10.2 L(ANR) per cycle

Drain piping: Rc 1/2

Product weight: 1,250 kg

Compatible air-cooling condenser: RCR-R10F

Air-cooling remote condenser waste heat: Max 33.5kW

Remote condenser weight: 135