For full functionality of this site it is necessary to enable JavaScript.
EMIN.MY
0
Product image

Chang AI CI-PS10 Portable biogas analyzer (0~25.00% O2, 0~99.9% CH4, 0~99.9% CO2, 0~10000ppm H2S)

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Oxygen (O2) detection: Electrochemical oxygen sensor

Scope of measurement: 0~25.00% O2

Accuracy: ±2%FS

Repeatability: ±1%FS

Stability: ±1%FS/7d

Response time: T90<60s

Methane (CH4) detection: NDIR Non-dispersive Infrared (Double Beam) Measurement

Scope of measurement: 0~99.9% CH4

Accuracy: ±2%FS

Repeatability: ±1%FS

Stability: ±1%FS/7d

Response time: T90<15s

Carbon dioxide (CO2) detection:

Sensors: NDIR Non-dispersive Infrared (Double Beam) Measurement

Scope of measurement: 0~99.9% CO2

Accuracy: ±2%FS

Repeatability: ±1%FS

Stability: ±1%FS/7d

Response time: T90<15s

Hydrogen sulfide (H2S) detection:

Sensors: Electrochemical hydrogen sulfide sensor

Scope of measurement: 0~10000ppm H2S

Accuracy: ±2%FS

Repeatability: ±1%FS

Stability: ±1%FS/7d

Response time: T90<60s

Analyzer:

Charging power supply: 85~264VAC,50/60Hz

Standby time: More than 8 hours (normal use)

Environmental temperature: 0~45°C

Measurements: Environmental humidity:Less than 80% RH

Sample flow rate: Gas intake

Intake pressure: 400~600ml/min

Inlet and outlet Atmospheric pressure ±10%(outlet should be atmospheric pressure)

connections: 1/4-inch NPT internal thread,stainless steel material

Analyzer weight: Net weight is 5.15kg and about 10kg after packing.

Cabinet size: L×W×H=305mm×270mm×144mm