For full functionality of this site it is necessary to enable JavaScript.
EMIN.MY
0
Product image

Anton Paar Nova 600 BET Surface Area Analyzer (2 stations; N2, Ar, CO2, and other non-corrosive gases)

ModelNova 600
P/N230147
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Measurement principle: Vacuum Volumetric

Analysis gases: N2, Ar, CO2, and other non-corrosive gases

Analysis stations: 2

Independent P0 station: Yes

Relative pressure range (P/P0 ): 10⁻⁴ to 0.999

Pressure measurement accuracy: 0.1 % (of full scale) *

Pressure resolution:

Absolute: 1.2 x 10⁻⁴ Torr

Torr Relative: 1.5 x 10⁻⁷ P/P0

Lower specific surface area limit: From 0.01 m²/g

Lower absolute surface area limit: From 0.5 m²

Surface area reproducibility:  2 %**

Pore size range: 0.35 - 500 nm (Diameter) (0.35 - 1.1 nm with CO2) (1.1 - 500 nm with N2)

Minimum pore value: 1.2x10⁻⁸ cm³

Analysis dewar:

Volume: 2 L

Duration: Up to 40 Hours

Dimensions (DxWxH): 44 cm x 63 cm x 84 cm

Weight: 63 kg

Operating environment: Temperature : 15 to 35 °C

Humidity : 20 to 80 % relative, non-condensing

Wetted parts: Stainless steel, Viton elastomers

Gas: Ports: 5 (3 Analysis, 1 Helium, 1 Degas/Backfill)

Purity: 99.999 % (He, N2)

Input Pressure: 8 to 10 PSIG

Vacuum connection: Rotary Pump Exhaust Port, KF 16

Vacuum requirements: Ultimate vacuum of 2.3 x 10⁻³ Torr

Electrical: Supply : AC 100~240 V AC , 50 / 60 Hz Consumption: 345 VA (Maximum)


* Includes precision, linearity, and hysteresis s of the complete pressure measurement system

** Performed with BAM P115 certified reference material

Brochure

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts