For full functionality of this site it is necessary to enable JavaScript.
EMIN.MY
0
Product image

AEP Transducers DC-TC4 Force Transducer (IP40; 100T)

ModelDC-TC4
P/NCDCTC4230100T5M
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Accuracy : ≤ ± 0.05% F.S.

Nominal Load : 100 t

Linearity: ≤ ±0.05% F.S.

Hysteresis: ≤ ±0.05% F.S.

Repeatability: ≤ ±0.02% F.S.

Internal Resolution: ±99999 div.

Readings Per Sec. (0 Filter): 100 (10ms)

Reference Temperature: +23°C

Temperature Nominal Range: -10/+40 °C

Service Temperature: -10/+70 °C

Storage Temperature: -20/+80 °C

Temperature Effect (10°C):

a) on zero: ≤ ±0.028% F.S.

b) on sensitivity: ≤ ±0.024% F.S.

Effect Of Transverse Load::

a) at 10% of nominal load: ≤ ±0.030% F.S.

Zero Function: 100% F.S.

Programmable Digital: 0÷5

Filter Programmable: 1, 2, 5, 10, 20, 50, 100

Resolution Peak Function: Pos. (+), Neg. (-) or Disabled

Red Display (High Efficiency): 7 segments

Display Height: 13 mm

Recommended Power Supply: 10÷15Vdc

Absolute Power Supply: 18Vdc overvoltage and overcurrent protected

Max. Electrical Input: 50mA

Power Connector: MIL PT02E10-6P

Mechanical Limit Values Referred To Nominal Load::

a) service load: 120%

b) max permissible load: 150%

c) breaking load: >300%

d) max transverse load: 50%

e) max permissible dynamic load: 75% (1)

Displacement At Nominal Load: ∼0.23mm

Peso Weight : ∼16.5 kg

Protection Class (En 60529): IP40

Execution Material: Stainless Steel

Cable Length: 5m

Fixing Screws:

a) diameter: M24

b) resistance class: 12.9

c) tightening torque: 460 Nm