For full functionality of this site it is necessary to enable JavaScript.
EMIN.MY
0
Product image

ACS CST 130/2T Air Thermal Shock Chamber

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Generalities

Basket capacity: 130 lt

Basket useful dimensions: 614x500x425 (WxDxH) mm approx

Maximum basket load: 50 Kg

Chamber external dimensions: 1252x2072x2844 (WxDxH) mm approx

Rated power: 15,1 kW

Rated current absorption: 28,1 A

Sound pressure level 60 dB(A) measured at 1m distance from the front of the unit, in 1,6m height, free field according to EN ISO 11201

Voltage: 400 V ±10%/50Hz/3ph + N + G

Weight: 1310 Kg

Performances (measured at room temperature of 22°C, empty working space and nominal voltage)

Temperature range:

upper chamber +70/+220°C

lower chamber -80/+100°C

Basket displacement speed <10 sec.

Temperature fluctuation: ±1°C

Temperature Rate of Change (measured at the control probe, according to IEC 60068-3-5):

Cold chamber:

heating: 2.0°C/min. from -80°C to +100°C

cooling: 4.7°C/min. from +100°C to -80°C

Hot chamber:

heating: 6.6°C/min. from +20°C to +220°C

Max load with MIL 883 test (15 min resetting time on specimen):

13 kg Ics distributed on 2 shelves with MIL 883 test D

20 kg Ics distributed on 2 shelves with MIL 883 test C

26 kg Ics distributed on 2 shelves with MIL 883 test B

Water consumption: 2,8 m3/h at a cooling water temperature of +29°C and temperature difference of 5°C (water temperature +12°C to +29°C)


Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts